MFS KITMulti Functional Sinus [KHA001]
- A comprehensive kit to approach direct & indirect maxillary sinus lift simply.
Crestal Approach - Components
1. Point Drill 800~1,000rpm
- Use to mark the point of perforation on cortical bone.
- In the case of the remaining bone height is as low as 3.5mm, pay more attention when drilling.
2. 2.2 Twist Drill 800~1,000rpm
- Use for making guide hole before using the Crestal Drill.
- Connect the Crestal Drill Stopper according to the height of the remaining bone.
3.Crestal Drill 400~800rpm
- Use the Crestal Drill sequentially according to the diameter of the fixture to be placed.
- Can also be used if sinus floor is flat, incline, septum.
- The Crestal Drill can be used about 50 times (depending on bone quality).
4.Crestal Drill Stopper
- Connected with a stopper to be drilled to the same length of
the cartilage height of maxillary sinus which is measured by CT. - If not equipped with CT, fasten the stopper one step
lower than expected and gradually increase the length.
5.Depth Gauge
- Measure thickness of the residual bone after checking the perforation of the cartilage of the maxillary sinus (do not open completely, only the entrance side should be opened).
- The stopper is attached to the base of the residual bone
to separate the cartilage and membrane from the maxillary sinus.
6.Aqua Membrane Lifter System
- After confirming that elevation of the cartilage of maxillary sinus,
elevate membrane with Aqua Membrane Lifter system.
- ① Fasten Aqua Lifter Drill to the drill hole.
- ② Connect Aqua Tube to syringe using Aqua Tube to Syringe using Syringe Connector (SC).
- ③ Inject saline solution equal to the amount of bone graft
material to be used for syringe. - ④ Tube connection to Aqua Lifter Drill using Ratchet Connector (RC).
- ⑤ Inject Saline solution.
- Aqua Lifter
- Aqua Ratchet Connector
- Aqua Syringe Connector
- Aqua Tube
7.Torque Wrench
- Use the Torque Wrench to fix Aqua Lifter Drill in the hole formed using Crestal Drill.
8. Bone Carrier
- Insert bone graft material using Bone Carrier.
9. Bone Condenser
- After connecting stopper with Bone Condenser,
elevate bone graft materials to inside of maxillary sinus. - Rotate bone graft material using Bone Condenser to disperse bone
graft material (possible to do it with Depth Gauge).
10. Implant Drill(Final)
- Drilling 1~2mm more deeply than steps of Crestal Drill.
11. Implant Placement
- If the residual bone is less than 3mm, do not implant the fixture, but bone graft only.
Crestal Approach - Drill Sequence
- Placing implant over Ø4.0 is highly recommended.
1. Ø3.3 Narrow Fixture
2. Ø3.5(Ø3.7) Fixture
3. Ø4.0 Fixture
4. Ø4.5 Fixture
- ※ Ø5.0 Fixture Normal Bone : Drilling with the Final Drill before placing implants are required.
- ※ Use a drill that is one step shorter than the implant (Ex. 10mm implant, 8 ~ 9mm Drill).
Lateral Approach - Components
1. Ø6 Lateral Reamer 800~1,000rpm
- Drill after fastening the stopper according to the height of the bone.
- Round shape to prevent membrane perforation.
2. Ø6 Lateral Round Drill 800~1,000rpm
- Drill after fastening the stopper according to the height of the bone.
- Round shaped edge.
- The residual bone should be replaced in the original position after drilling, sinus lifting & augmentation.
3. Lateral Stopper
4. Sinus Elevator
CSE-01 : Initial elevation of Sinus Membrane.
CSE-02 : as stepwise, after using CSE-01, used for elevation of sinus membrane.
CSE-03 : as stepwise, after using CSE-02, used for elevation of sinus membrane.
5. Ø4 Side Cutter 800~1,000rpm