MFR Kit | COWELL® SPECIAL KIT | 코웰메디 제품 | 코웰메디
SKIP NAVIGATION

COWELL® EXPERT INSTRUMENTS

MFR KITMulti Functional Removal [KHA001]

Easy Removal of fractured Fixture / Screw / Abutment

MFR Kit Composition - Middle Tray

MFR Kit Components

1. Fixture Removal System
  1. ① Connect the F/R Screw to the FRS Driver.
  2. ② Connect the F/R Screw mounted FRS Driver to the fixture (clockwise 40~60 N.cm) and remove the FRS Driver.
  3. ③ Connect the Fixture Remover to the F/R Screw (counterclockwise).
  4. ④ Remove the fixture after connecting the Torque Wrench (counterclockwise, 100~400 N.cm).
  5. ⑤ To remove the fixture from the Fixture Remover, use such device as vise to fix the Fixture Remover and connect to the Torque Wrench.
  6. ⑥ After connecting the FRS Driver to the F/R Screw, use the Torque Wrench to remove the F/R Screw (counterclockwise).
  • ※ One-time use of the F/R Screw is recommended (bending or fracture may happen if more than 100 N.cm and using twice may be possible if less than 100 N.cm).
  • ※ Sufficient Irrigation required when removing fixture.
  • ※ When the maximum torque is exceeded, the fixture may be bent or fractured.
  • ※ If fixture can not be removed even with maximum torque,
    remove Fixture Remover & F/R Screw, remove bones around fixture while using Round Bur and retry to remove.
2. Screw Removal System
Talon Drill
  1. ① Check the broken screw size inside the fixture
  2. ② Connect the I Guide corresponding to the fixture to the Guide Holder and fasten to the fixture.
  3. ③ Insert the Talon Drill into the I Guide Hole.
  4. ④ Remove the screw by pressing force (counterclockwise, 60 ~ 80 rpm).

※ If the I Guide and fixture could not be correctly connected, the path is not correct.

Reverse Drill & Screw Remover

If the screw could not be removed by the Talon Drill

  1. ① Form the hole on the fractured screw (depth 1 ~ 2mm / counterclockwise / 1,200 ~ 1,400 rpm).
  2. ② Use Screw Remover according to the created drill hole, remove the screw by pressing force (counterclockwise, 80 N.cm).
  • ※ If the path of the I Guide and fixture did not match, It would be difficult to remove the screw because the drill hole is away from the center of the screw.
  • ※ Reverse drilling requires removal of chips by irrigation & suction.
  • ※ The fractured screw may be removed during reverse drill hole creation.
  • ※ If necessary, fasten to the Machine Adapter and use the hand or Torque Wrench.

3. Abutment Removal System
Abutment Remover
  1. ① Used when 2 piece type abutment is fractured.
  2. ② Connect to the Abutment Remover (clockwise).
  3. ③ Connect correctly and wiggle to left and right side.
Tap Repair
  1. ① Used when the thread inside the fixture is occluded and damaged.
  2. ② Reproduce the thread using the Tap Repair.
Slot Driver
  1. ① Used for the damaged solid abutment (1 piece type), healing abutment and cover screw.
  2. ② Form a slot on the surface of the damaged abutment using round bur.
  3. ③ Connect the Slot Driver to the formed slot and remove it (counterclockwise).
4. Torque Wrench