AO4 Surgical Stent
[KDSS001]
- Guide the position of Implant and Drill.
Characteristic
- Guide the position of the implant and drill during implant placement.
- It improves the stability and accuracy in surgery, and it can shorten the time.
- By preventing the loss of healthy gums as much as possible, pre-fabricated prostheses can be placed immediately after surgery without the need for gum restoration.
- Angled line allows more precise and predictable surgery.
Eligible for
- A toothless patient.
- Patient who do not want long-period of surgery.
- Patients suffering from adult diseases such as hypertension and diabetes.
- Patients who need precise implant treatment.
Instruction
-
Place the AO4 Surgical Stent
-
- Make an incision for flap lift.
- Place the AO4 Surgical Stent using Ø2mm Twist Drill.* It is needed to check the position of mental foramen.
-
Place the INNO Fixture
-
- Drill with reference to the angled line and place the fixture.
-
Place the Multi S&A Abutment
-
- After placing the INNO fixture,
connect the Multi S&A Abutment according to the site.* Posterior site : Place the Multi A abutment (30°) with 30 N.cm torque force.
* Anterior site : Place the Multi A abutment (15°) or the Multi S abutment with
15 N.cm torque force (it is possible to allow emergence of the prosthetic screw).
-
or
-
Placement Lock Abutment
-
- After placing the INNO Fixture, connect Lock Abutment according to the site.
* If implant placement at an angle is not appropriate or not desired,
using the INNO Sub. Short Implant is highly recommended.